ANALISIS NILAI COMPUTED TOMOGRAPHY DOSE INDEX VOLUME (CTDIvol) DAN DOSE LENGTH PRODUCT (DLP) PADA PEMERIKSAAN MSCT KEPALA

Authors

  • Shakira Alwi. Universitas ‘Aisyiyah Yogyakarta
  • Ildsa Maulidya Marathus Nasokha Universitas ‘Aisyiyah Yogyakarta
  • Ike Ade Nur Liscyaningsih Universitas ‘Aisyiyah Yogyakarta
  • Arifin Musthafa RSUD dr. Soehadi Prijonegoro Sragen

Keywords:

CTDIvol, Head MSCT, DLP, DRL

Abstract

Background: MSCT (Multislice Computed Tomography) produces higher radiation doses compared to other imaging modalities. The radiation dose delivered to patients is measured using two main parameters: the Computed Tomography Dose Index Volume (CTDIvol) and the Dose Length Product (DLP). To protect patients from excessive radiation exposure, BAPETEN (Nuclear Energy Regulatory Agency) monitors and enforces optimization measures through the Diagnostic Reference Level (DRL). This study aims to determine the level of radiation dose received by patients during non-contrast head MSCT examinations. Methods: A mixed-methods approach was used, including observation, interviews, documentation, and literature review. Research instruments comprised interview guides, a smartphone camera for documentation, and writing materials. The study subjects included 30 patients and one medical physicist, while the object of study was the monitor console of a Siemens 128-slice CT scanner used for patient data collection. Data analysis was performed using SPSS with the Shapiro-Wilk test for normality, followed by a one-sample t-test at a 95% confidence level to test the hypothesis. Results: The mean CTDIvol was 49.58 ± 0.04 mGy (range 49.5–49.6), and the mean DLP was 1,024 ± 257.28 mGy·cm (range 144–1,418). These values correspond to 82.63% of the CTDIvol limit and 80.31% of the DLP limit based on the national DRL (60 mGy; 1,275 mGy·cm). Variations in DLP among patients were due to differences in scan coverage length and repeated scans caused by patient movement, particularly in post-traffic accident cases with decreased consciousness. The one-sample ttest indicated a significant difference compared to the national DRL values (p < 0.05). Conclusion: Non-contrast head MSCT examinations at RSUD (Regional General Hospital) dr. Soehadi Prijonegoro Sragen comply with radiation protection optimization principles. The CTDIvol and DLP values obtained were below the national DRL limits set by BAPETEN, indicating that the scanning protocol is safe. Regular dose evaluation and the use of immobilization devices for uncooperative patients are recommended to minimize dose variability.

References

Andriole, P. K. (2023). Picture archiving and communication systems: past, present, and future. Journal Of Medical Imaging, 10(6), 1–9.

BAPETEN. (2021). Kepala Badan Pengawas Tenaga Nuklir Republik Indonesia.

Bos, D., Guberina, N., Zensen, S., Opitz, M., Forsting, M., & Wetter, A. (2023). Radiation Exposure in Computed Tomography. National Library Of Medicine, 9, 134–141.

Brenner, J. D., & Hall, J. E. (2023). Computed Tomography--an Increasing Source of Radiation Exposure. National Library of Medicine.

Damilakis, & john. (2021). CT Dosimetry What Has Been Achieved and What Remains to Be Done. Investigative Radiology.

Fiarka, F., Oktamuliani, S., & Nuraeni, N. (2023). Pengaruh Tegangan terhadap Nilai CTDIvol, Dose Length Product dan Dosis Efektif pada Pemeriksaan Computed Tomography (CT) Abdomen: Studi di Rumah Sakit Umum Pusat Dr. M. Djamil. Jurnal Fisika Unand, 12(4), 591–597. https://doi.org/10.25077/jfu.12.4.591-597.2023

Hermena, S., & Young, M. (2023). CT-scan Image Production Procedures. National Library Of Medicine.

Hoeijmakers, i, j E., Martens Bibi, Wildberger, E. J., Flohr, G. T., & Jeukens, N. P. L. R. C. (2025). Objective assessment of diagnostic image quality in CT scans: what radiologists and researchers need to know. Springer Open .

ICRP Annual Report. (2017). Annals of The ICRP. ICRP 2017 Annual Report.

Kunarsih, E., Sudradjat, & Pratama Putra Gede, B. I. (2021). Pedoman Teknis Penerapan Tingkat Panduan Diagnostik Indonesia (Indonesia Diagnostic Reference Level).

Lampignano, P. J., & Kendrick, E. L. 2018. TEXBOOK OF Radiographic Positioning and Related Anatomy (ninth).

Lell, M., & Kachelrieß, M. (2023). Computed Tomography 2.0 New Detector Technology, AI, and Other Developments. Investigative Radiology, 58(8), 587–601.

Masuku S, R. (2023). Analisis nilai Computer Tomography Dose Index (CTDI) dan Dose Length Product (DLP)

Mateus Lima, B. G., Silva Torres-da, R. K., Machado Rodrigues Silva, da R. A., Almeida de Falipe, P., & Silva da Valério, A. (2021). Morphological and morphometric study of Foramen Magno and Occipital Condits in dry skulls of Mato Grosso do Sul. Research,Society and Development, 10(12).

McCollough, C. H., Boedeker, K., Cody, D., Duan, X., Flohr, T., Halliburton, S. S., Hsieh, J., Layman, R. R., & Pelc, N. J. (2020). Principles and applications of multienergy CT: Report of AAPM Task Group 291. In Medical Physics (Vol. 47, Issue 7, pp. e881–e912). John Wiley and Sons Ltd. https://doi.org/10.1002/mp.14157

Muqmiroh, Setyanur, Maimanah, & Witjaksono. (2018). The Effect of Window Width and Window-level Settings in Non-enhanced Head CT to Increase the Diagnostic Value of Subacute Ischemic Stroke. Social Sciens.

Naser-Al, A. Y., & Tafti Dawood. (2023). CT Instrumentation and Physics. National Library Of Medicine.

Ni, N., Ahadiyah, matin, Setiawati, E., Arifin, Z., & Choirul Anam, dan. (2020). Pengukuran Computed Tomography Dose Index (Ctdi) Pada Fantom Kepala Dengan Menggunakan Ct Dose Profiler (Vol. 23, Issue 4).

Paulo, G., Damilakis, J., Tsapaki, V., Schegerer, A. A., Repussard, J., Jaschke, W., & Frija, G. (2020). Diagnostic Reference Levels based on clinical indications in computed tomography: a literature review. National Library of Mediicine.

Perba No.4 2020. (2020). Kepala Badan Pengawas Tenaga Nuklir Republik Indonesia.

Pina, & Von den Hoff. (2024). Skull and Craniofacial Development and Regeneration. Frontiers in Physiology, 15.

Seeram, E. (2016). Computed Tomography Physical Principles, Clinical Applications, and Quality Control (Fourth Edition). http://evolve.elsevier.com/Seeram/You’vejustpurchased

Seeram, E. (2016). Computed Tomography: Physical Principles, Clinical Applications, And Quality Control, Fourth Edition. http://evolve.elsevier.com/Seeram/You’vejustpurchased

Susam, L. A., Yilmaz, A., ALMisned, G., Yilmaz Alan, H., Ozturk, G., Kilic, G., Tuysuz, B., Topuzlar, S. E., Akkus, B., Ene, A., & Tekin, H. O. (2023). Tailoring a Behavioral Symmetry on KERMA, Mass Stopping Power and Projected Range Parameters against Heavy-Charged Particles in Zinc-Tellurite Glasses for Nuclear Applications. Symmetry, 15(6). https://doi.org/10.3390/sym15061201

Susanto, W. (2023). Penetuan Nilai Diagnostic Reference Level (DRL) CT-Scan Untuk Pemeriksaan Kepala dan Dada Pasien Dewasa. Badan Pengawas Tenaga Nuklir-BAPETEN, 1–6.

Szczykutowicz, T. P., Toia, G. V., Dhanantwari, A., & Nett, B. (2022). A Review of Deep Learning CT Reconstruction: Concepts, Limitations, and Promise in Clinical Practice. In Current Radiology Reports (Vol. 10, Issue 9, pp. 101–115). Springer. https://doi.org/10.1007/s40134-022-00399-5

Vano, E., Frija, G., Loose, R., Paulo, G., Efstathopoulos, E., Granata, C., & Andersson, J. (2021). Dosimetric quantities and effective dose in medical imaging: a summary for medical doctors. In Insights into Imaging (Vol. 12, Issue 1). Springer Science and Business Media Deutschland GmbH. https://doi.org/10.1186/s13244-021-01041-2

Downloads

Published

2025-11-15

How to Cite

Shakira Alwi., Ildsa Maulidya Marathus Nasokha, Ike Ade Nur Liscyaningsih, & Arifin Musthafa. (2025). ANALISIS NILAI COMPUTED TOMOGRAPHY DOSE INDEX VOLUME (CTDIvol) DAN DOSE LENGTH PRODUCT (DLP) PADA PEMERIKSAAN MSCT KEPALA. Journal of Innovation Research and Knowledge, 5(6), 7245–7270. Retrieved from https://bajangjournal.com/index.php/JIRK/article/view/11739